For more info: sales@tangkasanugerah.com
Features
PCS-1000I high precision AC and DC shunt meter can simultaneously measure current and voltage with the maximum 6 1/2 measurement resolution.
GW Instek rolls out the new PCS-1000I isolated output high precision current shunt meter, which inherits the simultaneous voltage and current measurement function of PCS-1000. PCS-1000I adopts five sets of independent shunt resistors to provide five current measurement levels, including300A,30A,3A, 300mA, and 30mA to meet the requirements of different current level measurements. Internally, PCS-1000I utilizes two sets of 24bits ADCs and low temperature coefficient electronic components to mainly focus on the current measurement of power supply devices. High precision PCS-1000I can be used in adjusting and calibrating instruments. Additionally, temperature variation will not cause PCS-1000I to yield any measurement errors. PCS-1000I can automatically select optimal measurement level with the maximum resolution so as to replace manual selection to save operational time.
PCS-1000I provides a BNC output, which can connect with an oscilloscope to directly observe current waveform variation without using a current probe. General oscilloscopes do not have isolated channels and their input and output are structured at a common point, therefore, the output load will likely result in measurement errors. PCS-1000I’s isolated current output design can prevent measurement errors from an oscilloscope with non-isolated outputs. PCS-1000I, a high precision AC/DC current shunt meter, not only provides USB and GPIB communications interfaces for users to remotely control the instrument but also conducts simultaneous voltage and current measurements. The SCPI communications commands of PCS-1000I allow users to remotely control PCS-1000I via a PC to operate measurement data readbacks.
PCS-1000I will replace non-isolated output PCS-1000 models, which will be discontinued.
PCS-1000 I Specifications |
|||
General |
|||
Power Supply |
100 V / 120 V / 220 V / 240 V ±10% |
||
Power Line Frequency |
50/60 Hz |
||
Operating Environment |
Full accuracy for0 ℃to55 ℃, Full accuracy to 80% R.H. at40 ℃ |
||
Storage Environment |
-40 ℃to70 ℃ |
||
Power Consumption |
Max 35VA |
||
Dimensions |
210 mm(W) *80 mm(H) *390mm(D) |
||
Weight |
Approxiately5 kg |
||
DC Characteristics |
|||
DC Voltage |
|||
Range |
1 Year23 ℃±5 ℃ |
Temperature |
|
200.0000 mV |
|
0.0050 + 0.0035 |
0.0005 + 0.0005 |
2.000000 V |
|
0.0050 + 0.0010 |
0.0005 + 0.0001 |
20.00000 V |
|
0.0050 + 0.0010 |
0.0005 + 0.0001 |
200.0000 V |
|
0.0050 + 0.0010 |
0.0005 + 0.0001 |
1000.000 V |
|
0.0050 + 0.0020 |
0.0005 + 0.0001 |
Accuracy specification : ± (% of reading + % of range) |
|||
Voltage input Resistance:10MΩ for all DC Voltage ranges |
|||
|
|||
DC Current |
|||
Range |
Burden Voltage |
1 Year23 ℃±5 ℃ |
Temperature |
30.00000 mA |
<0.4 V |
0.01 + 0.005 |
0.001 + 0.002 |
300.0000 mA |
<0.5 V |
0.01 + 0.005 |
0.001 + 0.002 |
3.000000 A |
<0.8 V |
0.01 + 0.005 |
0.001 + 0.002 |
30.00000 A*1 |
<0.8 V |
0.01 + 0.005 |
0.001 + 0.002 |
300.0000 A*1 |
<0.8 V |
0.02 + 0.005 |
0.001 + 0.002 |
Accuracy specification : ± (% of reading + % of range) |
|||
|
|||
Isolated DC Current Monitor Accuracy |
|||
Range |
Resolution( 6 1/2) |
DC Accuracy |
Temperature |
30.00000 mA |
0.00001mA |
0.1+0.05 |
0.001 |
300.0000 mA |
0.0001mA |
0.1+0.05 |
0.001 |
3.000000 A |
0.000001A |
0.1+0.05 |
0.001 |
30.00000 A*1 |
0.00001A |
0.1+0.05 |
0.001 |
300.0000 A*1 |
0.0001A |
0.2+0.05 |
0.001 |
Accuracy specification : ± (% of output + % of full scale) |
|||
Monitor output voltage for the full scale current = 3V |
|||
|
|||
AC Characteristics |
|||
|
|||
True RMS AC Voltage |
|||
Range |
Frequency |
1 Year23 ℃±5 ℃ |
Temperature |
200.0000 mV |
45Hz - 2kHz |
0.5 + 0.05 |
0.005 + 0.005 |
2.000000 V |
0.005 + 0.005 |
||
20.00000 V |
0.005 + 0.005 |
||
200.0000 V |
0.005 + 0.005 |
||
600.000 V |
0.005 + 0.005 |
||
Accuracy specification : ± (% of reading + % of range) |
|||
|
|||
True RMS AC Current |
|||
Range |
Frequency |
1 Year23 ℃±5 ℃ |
Temperature |
30.00000 mA |
45Hz - 2kHz |
0.5 + 0.05 |
0.03 + 0.006 |
300.0000 mA |
0.03 + 0.006 |
||
3.000000 A |
0.03 + 0.006 |
||
30.00000 A*1 |
45Hz - 400Hz |
0.5 + 0.05 |
0.03 + 0.006 |
300.0000 A*1 |
0.03 + 0.006 |
||
Accuracy specification : ± (% of reading + % of range) |
|||
|
|||
Isolated AC Current Monitor Accuracy |
|||
Range |
Frequency |
AC Accuracy |
Max Input Limit |
30.00000 mA |
45Hz - 2kHz |
0.2 + 0.05 |
30 mA |
300.0000 mA |
300 mA |
||
3.000000 A |
3 A |
||
30.00000 A*1 |
45Hz - 400Hz |
0.5 + 0.05 |
30 A |
300.0000 A*1 |
300 A |
||
Accuracy specification : ± (% of output + % of full scale) |
|||
Monitor output voltage for the full scale current = 3V |
|||
The specifications are only applicable when the input is 10% or greater of the full scale range |
(The specifications apply when the PCS-1000I is powered on for at least 30 minutes to warm-up to a temperature of18 ℃to28 ℃, unless specified otherwise.)
*1 The accuracy for30A/300Alevels must be increased by a power factor of 8 ppm/watt.