FEATURES:
- 6 1/2 Digit measurement resolution
- Dual measurement for Voltage and Current
- AC&DC Current measurement range: 30mA/300mA/3A/30A/300A(Max)
- AC Voltage measurement range: 200mV/2V/20V/200V/600V(Max)
- DC Voltage measurement range: 200mV/2V/20V/200V/1000V(Max)
- Auto zero circuit
- Standard Interfaces: USB, GPIB

DESCRIPTION

PCS-1000 is a High-Precision D.C. and A.C. Current Shunt Meter which carries built-in current shunts and high-accuracy current measurement circuits. PCS-1000 not only measures current in five current levels ?300A, 30A, 3A, 300mA and 30mA, but also measurePCS-1000 is a High-Precision D.C. and A.C. Current Shunt Meter which carries built-in current shunts and high-accuracy current measurement circuits. PCS-1000 not only measures current in five current levels -300A, 30A, 3A, 300mA and 30mA, but also measures voltage at 6 1/2-digit resolution simultaneously. This unique feature puts PCS-1000 in an unprecedented position among all the shunt meters available in the market today. After a long time operation, PCS-1000 is able to maintain high measurement accuracy in spite of the temperature variation. This is due to its core design based on the high stability and low temperature coefficient shunts. PCS-1000, a verification-lab level current and voltage measurement equipment, is suitable for the R&D, quality verification and testing jobs of all power products such as power supply, communications system power, server power and electronic load. The maximum voltage/current measurement ranges of PCS-1000, including DC 1000V / AC 600V / AC 300A / DC 300A, meet all current and voltage measurement challenges of every power industry.

FEATURES

Model PCS-1000    
 DC Voltage Characteristics    
    23℃±5℃(1 Year) Temperature Cofficient / ℃
 Range 200.0000mV 0.0050+0.0035 0.0005+0.0005
2.000000V 0.0050+0.0010 0.0005+0.0001
20.00000V 0.0050+0.0010 0.0005+0.0001
200.0000V 0.0050+0.0010 0.0005+0.0001
1000.000V 0.0050+0.0020 0.0005+0.0001
 Accuracy     ±(% of reading +% of Range); Voltage input Resistance:10MΩ for all DC Voltage ranges
 DC Current Characteristics    
    Burden Voltage 23℃±5℃(1 Year) Temperature Cofficient / ℃
 Range 30.00000mA <0.4V 0.01+0.005 0.001+0.002
300.0000mA <0.5V 0.01+0.005 0.001+0.002
3.000000A <0.8V 0.01+0.005 0.001+0.002
30.00000A <0.8V 0.01+0.005 0.001+0.002
300.0000A <0.8V 0.02+0.005 0.001+0.002
 Accuracy ±(% of reading +% of Range) 
 True RMS AC Voltage Characteristics    
    Frequency 23℃±5℃(1 Year) Temperature Cofficient / ℃
 Range     200.0000mV 45Hz-2KHz
2KHz-10kHz
10kHz-20kHz
0.5+0.05
1.0+0.05
2.0+0.10
0.005+0.005
2.000000V 0.005+0.005
20.00000V 0.005+0.005
200.0000V 0.005+0.005
600.000V 0.005+0.005
 Accuracy ±(% of reading +% of Range) ℃    
 True RMS AC Current(AC+DC) Characteristics    
     Range         30.00000mA    45Hz-2KHz
2KHz-10kHz
   0.5+0.05
1.0+0.05
0.03+0.006
300.0000mA 0.03+0.006
3.000000A 0.03+0.006
30.00000A   45Hz-400Hz   0.5+0.05 0.03+0.006
300.0000A 0.03+0.006
 Accuracy ±(% of reading +% of Range)     
 Current Monitor Accuracy    
  Shunt Value DC Accuracy AC Accuracy ≦ 400Hz Max Input DC/ AC RMS
 Range     10Ω 0.01% 0.1% 30mA
0.01% 0.1% 300mA
0.1Ω 0.01% 0.1% 3A
0.01Ω 0.01% 0.1% 30A
0.001Ω 0.02% 0.1% 300A
 General     
 Power Supply      100V/120V/220V/240V±10%
 Power Line Frequency     50/60Hz
 Operating Environment Full Accuracy for 0℃ to 50℃, Full accuracy to 80% R.H. at 40℃    
 Storage Environment  -40℃ to 70℃   
 Power Consumprion  Max 35VA   
 Dimensions 210 mm (W) x 80mm(H) x 390 mm(D)    
 Weight  Approx 6kg   