FEATURES:
- 100VA AC Test Capacity
- 240x48 Ice Blue Dot matrix LCD
- Auto W-I or I-W Mode
- ARC Detection
- True RMS Current Measurement
- Zero Crossing Turn-on Operation
- Safety INTERLOCK function
- PWM amplifier to enhance the power efficiency and reliable testing
- Automatically Switching Input Source for Universal Input Voltage
- Interface : Remote terminal and Signal I/O

DESCRIPTION

GW Instek launches new economical safety testers (hi-pot tester/hipot tester), the GPT-9600 Series, which offers an affordable solution for supporting routine tests of major items of the safety standards such as IEC, EN, UL, CSA, GB, JIS and other safety regulations.

The GPT-9600 Series is built upon a platform of 100VA AC maximum power output. The GPT-9603 is a 3-in-1 model capable of performing AC withstanding, DC withstanding and insulation resistance tests. The GPT-9612 is capable of performing AC withstanding and insulation resistance tests. The GPT-9602 is capable of performing AC and DC withstanding tests, and GPT-9601 is able to perform AC withstanding test. The GPT-9600 Series is equipped with the high-efficiency PWM amplifier, which is the core of the platform design to impede the influence from the input AC voltage fluctuation and ensure a stable voltage output.

Following a tidy and easy-to–use design concept, the GPT-9600 Series renders users an intuitive operation environment by a simple and clear panel layout, a large LCD display and color LED indicators. The switching power supply, used as a universal input source, accommodates the power systems in most countries in the world. The GPT-9600 series, equipped with the same output voltage function as that of all GW Instek Safety Testers, indicates the expected output voltage before high voltage tests are applied. Furthermore, an AUTO mode, including test sequence selections of withstanding-then-insulation or insulation-then-withstanding, is designed for models carrying insulation Resistance test function to reduce the testing time of dual test items.

Other functions and features of GPT-9600 include: the zero crossing turn-on operation protects DUT from the impact of surge voltage output, the interlock function safeguards users from the hazardous shock of unintentional touch of the voltage output, a remote output on-off terminal in the front panel and a signal I/O port in the rear panel are provided as the means for remote start/stop control of the safety tester (hi-pot tester/hipot tester).

FEATURES

AC WITHSTANDING

Output-Voltage Range 0.10kV~ 5.00kV ac
Output-Voltage Resolution 10V
Output-Voltage Accuracy ± (1.5% of setting + 2 counts) with no load
Maximum Rated Load 100 VA (5kV/20mA)
Maximum Rated Current 20mA (0.5kV < V ≦ 5kV) 5mA (0.1kV ≦ V≦ 0.5kV)
Output-Voltage Waveform Sine wave
Output-Voltage Frequency 50 Hz / 60 Hz selectable
Voltage regulation ± (1.5% + 2 counts) [full load → no load]
Voltmeter Accuracy ± (1.5% of reading + 2 counts)
Current Measurement Range 0.01mA~20.0mA
Current Best Resolution 0.01mA / 0.1mA
Current Measurement Accuracy ±(2.0% of reading + 5 counts) when HISET < 1.00mA ±(2.0% of reading + 3 counts) when HISET ≧ 1.00mA
Current Judgment Accuracy ±(3.0% of setting + 5 counts) when HISET < 1.00mA ±(3.0% of setting + 3 counts) when HISET ≧ 1.00mA
Window Comparator Method Yes
ARC Detect Yes
RAMP (Ramp-Up Time) 0.1s fixed
TIMER (Test Time) OFF, 1s~180s
GND ON

DC WITHSTANDING

Output-Voltage Range 0.10kV~ 6.00kV dc
Output-Voltage Resolution 10V
Output-Voltage Accuracy ± (1.5% of setting + 2 counts) with no load
Maximum Rated Load 25W (5kV/5mA)
Maximum Rated Current 6mA (0.5kV < V ≦ 6kV) 2mA (0.1kV ≦ V≦ 0.5kV)
Voltage regulation ± (1.5% + 2 counts) [full load → no load]
Voltmeter Accuracy ± (1.5% of reading + 2 counts)
Current Measurement Range 0.01mA~6.00mA
Current Best Resolution 0.01mA
Current Measurement Accuracy ±(2.0% of reading + 5 counts) when HISET < 1.00mA ±(2.0% of reading + 3 counts) when HISET ≧ 1.00mA
Current Judgment Accuracy ±(3.0% of setting + 5 counts) when HISET < 1.00mA ±(3.0% of setting + 3 counts) when HISET ≧ 1.00mA
Window Comparator Method Yes
ARC Detect Yes
RAMP (Ramp-Up Time) 0.1s fixed
TIMER (Test Time) OFF, 1s~180s
GND ON

INSULATION RESISITANCE

Output-Voltage 50V, 100V, 250V, 500V, 1000V dc
Output-Voltage Accuracy ± (3.0% of setting + 1 count) [no load]
Resistance Measurement Range 1M ohm~ 2000M ohm
  Test voltage Measurable Range Accuracy
  50V/100V/250V 1~ 50M ohm 51~ 2000M ohm ±(5% of reading + 2M ohm) ±(10% of reading + 2M ohm)
  500V/1000V 1~ 500M ohm 501~ 2000M ohm ±(5% of reading + 2M ohm) ±(10% of reading + 2M ohm)
Window Comparator Method Yes
Output Impedance 600k ohm
RAMP (Ramp-Up Time) 0.1s fixed
TIMER (Test Time) OFF, 1s~180s
GND OFF (fixed)

TEST MODE*

SINGLE ACW, DCW, IR
AUTO AC-IR, IR-AC, DC-IR, IR-DC

INTERFACE

Remote Terminal (Front) Standard
SIGNAL IO Standard

General

DISPLY 240 x 48 Ice Blue dot matrix LCD
POWER SOURCE AC 100V-120V/220V-240V ±10%, 50/60Hz
POWER COMSUMPTION 400VA max.
DIMENSIONS & WEIGHT 330(W) x 148(H) x 385(D) mm approx.9kg(Max)