GW Instek GPM-8213 power meter is designed specifically for single-phase (1P/2W) AC power supply’s power measurements. Powerful features, including 4” TFT LCD, five-digit measurement display, 19 power measurement parameters, integral measurement function, high-accuracy voltage/current/power measurement capabilities, front/rear panel input terminals, and various communications ports, are to facilitate users with clear, convenient, and accurate power measurements.
GPM-8213 provides as many as 19 power measurement parameters, including voltage (Vrms/ V+pk / V-pk), current (Irms/ I+pk / I-pk), frequency (VHz/ IHz), power (P/ P+pk / P-pk), crest factor (CFV/ CFI), apparent power (VA), reactive power (VAR), power factor (PF), phase angle (DEG), total harmonic distortion (THDV/ THDI), high-accuracy voltage/current/power measurement capabilities (reading: ±0.1%; level: ±0.1%). The advantages of TFT LCD have been efficiently deployed to simple mode and standard mode. Simple mode displays conventional power meter’s four measurement parameters to meet the requirement of accuracy and clarity for the test on manufacturing process. Standard mode extends the display to the maximum of 8 measurement parameters (2 major measurements + 6 monitor measurements) to satisfy the various measurement application requirements of R&D, design, and quality verification.
For DUT requiring IEC 62301/EN 50564 standby power consumption test, GPM-8213 provides the optimal measurement supports, including test frequency bandwidth of DC~6kHz, the minimum current level of 5mA (resolution: 0.1uA), power measurement resolutions (1uW for minimum current and voltage levels; 1mW for maximum current and voltage levels), crest factor reaching 3 (half range reaching 6), and measurement of total harmonic distortion (at least 13th order power harmonic). For large voltage /large current measurement applications of general power measurement, GPM-8213 provides PT / CT rate function to collocate with external potential transformer or current transformer to meet the measurement requirements.
With respect to data retrieval and storage, the standard RS-232C/ USB interfaces (virtual COM)/ LAN can be utilized to edit and retrieve programs or the optional GPIB interface (installed by manufacturer) can be selected to meet users’ automatic test system requirements.
Note |
●The specifications apply when the GPM is warmed up for at least 30 minutes. ●Below are the basic conditions required to operate the GPM within specifications:
|
Specification Condition |
Temperature: 23˚C±5˚C Humidity: <80%RH(non-condensing) |
Operation Condition |
Temperature0°C~ 40°C,
|
Storage Condition |
Temperature −40°C ~70°C Humidity < 90%RH(non-condensing) |
Power Source |
AC 100-240V, 50–60Hz ; Consumption Max. 25VA |
Dimension & Weight |
270(W) x 110(H) x 350(D) mm, Approx. 2.9kg |
Item |
Spec. |
||
Input voltage |
600 Vrms |
||
Input current |
20 Arms |
||
Input impedance(50/60 Hz) |
Voltage |
2.4MΩ |
|
Current |
5mA - 200mA |
500mΩ |
|
0.5A - 20A |
5mΩ |
||
Maximum display voltage |
700 Vrms |
||
Maximum display current |
25 Arms |
||
Maximum allowable isolation voltage |
300 V |
||
Low frequency filter |
Cut-off frequency |
500 Hz |
Display
Synchronization frequency |
45Hz~ 6kHz |
Average |
1, 2, 4, 8, 16, 32, 64 |
Displayed items(Standard mode) |
8 items simultaneously. |
Displayed items(Simple mode) |
4 items simultaneously. |
Displayed digits |
5 |
Voltage converter (PT ratio) |
1 to 9999.999 |
Current converter (CT ratio) |
1 to 9999.999 |
Measurement items |
Voltage, current, active power, apparent power, reactive power, power factor, phase angle, frequency, integrated current, integrated power, positive integrated power, negative integrated power, integration time, voltage crest factor, current crest factor, voltage peak, current peak, Thd |
Displayed measurement parameters |
Vdc, Vrms, V+pk, V-pk, Idc, Irms, I+pk, I-pk, P, P+pk, P-pk, VA, VAR, PF, CFV, CFI, DEG, VHz, IHz, THDV, THDI |
Voltage Measurement
Measurement range |
CF=3 : |
15V, 30V, 60V, 150V, 300V, 600V |
|
CF=6 : |
7.5V, 15V, 30V, 75V, 150V, 300V |
||
Crest factor |
3, 6 |
||
Accuracy |
Effective range |
1 % to 105 % of range |
|
DC |
±(0.2 % reading + 0.2 % range) |
||
45 Hz ≤ f ≤ 66 Hz |
±(0.1 % reading + 0.1 % range) |
||
66 Hz < f ≤ 1kHz |
±(0.1 % reading + 0.2 % range) |
||
1 kHz < f ≤ 6kHz |
± 3 % of range |
||
The filter is turned on |
Increase 0.3 % reading@ 45Hz to 66Hz |
||
Temperature effect |
5-18˚C / 28-40˚C |
Increase ±0.03% reading /˚C |
|
Residual noise |
0.5 % of range |
Current Measurement
Measurement range |
CF=3 : |
5mA, 10mA, 20mA, 50mA, 100mA, 200mA, 500mA, 1A, 2A, 5A, 10A, 20A |
|
CF=6 : |
2.5mA, 5mA, 10mA, 25mA, 50mA, 100mA, 250mA, 0.5A, 1A, 2.5A, 5A, 10A |
||
Crest factor |
3, 6 |
||
Accuracy |
Effective range |
1 % to 105 % of range |
|
DC |
±(0.2 % reading + 0.2 % range) |
||
45 Hz ≤ f ≤ 66 Hz |
±(0.1 % reading + 0.1 % range) |
||
66 Hz < f ≤ 1kHz |
±(0.1 % reading + 0.2 % range) |
||
1 kHz < f ≤ 6kHz |
± 3 % of range |
||
The filter is turned on |
Increase 0.3 % reading@ 45Hz to 66Hz |
||
Temperature effect |
5-18˚C / 28-40˚C |
Increase ±0.03% reading /˚C |
|
Residual noise |
0.5 % of range |
Power Measurement
Accuracy |
Effective range |
1 % to 110 % of range |
DC |
±(0.2 % reading + 0.2 % range) |
|
45 Hz ≤ f ≤ 66 Hz |
±(0.1 % reading + 0.1 % range) |
|
66 Hz < f ≤ 1kHz |
±(0.1 % reading + 0.3 % range) |
|
1 kHz < f ≤ 6kHz |
± 3 % of range |
|
The filter is turned on |
Increase 0.3 % reading@ 45Hz to 66Hz |
|
Temperature effect |
5-18˚C / 28-40˚C |
Increase ±0.03% reading /˚C |
Frequency Measurement
Measurement range |
The filter is turned on |
30.000Hz to 499.99Hz |
The filter is turned off |
30.000Hz to 9.9999kHz |
|
Measurement items |
Voltage, Current |
|
Effective input range |
10% to 105% of voltage input range |
|
Accuracy |
±(0.06 % reading) |
Integrator Measurement
Integrator |
Accuracy |
±(Accuracy of voltage or current+ 0.1 % reading) |
Time |
Range |
0 hour 0 minute to 9999 hours 59 minutes |
Accuracy |
±0.01% ±1second |