Features

  • 4” TFT LCD
  • Two data display modes
  • Standard : two major measurement items + six secondary measurement items
  • Simple : displaying test data of four different measurement items
  • Met the requirement of IEC 62301 power measurement
  • Voltage/ current test frequency bandwidth : DC~6kHz
  • Watt resolution : 1mW
  • Lowest current range : 5mA, resolution : 0.1uA
  • Current/voltage measurements reach CF=3 for distorted wave and CF=6 for half range
  • W-h (power vs time)/A-h (current vs time) integration function
  • Total harmonic distortion measurement
  • Front panel test terminal
  • Standard interface : RS-232C, USB device, LAN
  • Optional interface : GPIB
  • Optional test fixture : GPM-001

DESCRIPTION

GW Instek GPM-8213 power meter is designed specifically for single-phase (1P/2W) AC power supply’s power measurements. Powerful features, including 4” TFT LCD, five-digit measurement display, 19 power measurement parameters, integral measurement function, high-accuracy voltage/current/power measurement capabilities, front/rear panel input terminals, and various communications ports, are to facilitate users with clear, convenient, and accurate power measurements.

 

GPM-8213 provides as many as 19 power measurement parameters, including voltage (Vrms/ V+pk / V-pk), current (Irms/ I+pk / I-pk), frequency (VHz/ IHz), power (P/ P+pk / P-pk), crest factor (CFV/ CFI), apparent power (VA), reactive power (VAR), power factor (PF), phase angle (DEG), total harmonic distortion (THDV/ THDI), high-accuracy voltage/current/power measurement capabilities (reading: ±0.1%; level: ±0.1%). The advantages of TFT LCD have been efficiently deployed to simple mode and standard mode. Simple mode displays conventional power meter’s four measurement parameters to meet the requirement of accuracy and clarity for the test on manufacturing process. Standard mode extends the display to the maximum of 8 measurement parameters (2 major measurements + 6 monitor measurements) to satisfy the various measurement application requirements of R&D, design, and quality verification.

 

For DUT requiring IEC 62301/EN 50564 standby power consumption test, GPM-8213 provides the optimal measurement supports, including test frequency bandwidth of DC~6kHz, the minimum current level of 5mA (resolution: 0.1uA), power measurement resolutions (1uW for minimum current and voltage levels; 1mW for maximum current and voltage levels), crest factor reaching 3 (half range reaching 6), and measurement of total harmonic distortion (at least 13th order power harmonic). For large voltage /large current measurement applications of general power measurement, GPM-8213 provides PT / CT rate function to collocate with external potential transformer or current transformer to meet the measurement requirements.

 

With respect to data retrieval and storage, the standard RS-232C/ USB interfaces (virtual COM)/ LAN can be utilized to edit and retrieve programs or the optional GPIB interface (installed by manufacturer) can be selected to meet users’ automatic test system requirements.

FEATURES

General

Note

●The specifications apply when the GPM is warmed up for at least 30 minutes.

●Below are the basic conditions required to operate the GPM within specifications:

  • 1-year calibration cycle
  • All specification are ensured under SLOW speed
  • Accuracy: ± (%of reading+ %of range)
  • The power cord protective grounding conductor must be connected to ground
  • Input voltage/current should be sinewave
  • Power factor is 1
  • Crest factor is 3
  • Common mode voltage is 0

Specification Condition

Temperature: 23˚C±5˚C

Humidity: <80%RH(non-condensing)

Operation Condition

Temperature0°C~ 40°C,

  • 30 ~ 40°C, Relative Humidity < 70%RH(non-condensing)
  • >40°C, Relative Humidity < 50%RH(non-condensing)

Storage Condition

Temperature −40°C ~70°C

Humidity < 90%RH(non-condensing)

Power Source

AC 100-240V, 50–60Hz ; Consumption Max. 25VA

Dimension & Weight

270(W) x 110(H) x 350(D) mm, Approx. 2.9kg

Input

Item

Spec.

Input voltage

600 Vrms

Input current

20 Arms

Input impedance(50/60 Hz)

Voltage

2.4MΩ

Current

5mA - 200mA

500mΩ

0.5A - 20A

5mΩ

Maximum display voltage

700 Vrms

Maximum display current

25 Arms

Maximum allowable isolation voltage

300 V

Low frequency filter

Cut-off frequency

500 Hz

Display

Synchronization frequency

45Hz~ 6kHz

Average

1, 2, 4, 8, 16, 32, 64

Displayed items(Standard mode)

8 items simultaneously.

Displayed items(Simple mode)

4 items simultaneously.

Displayed digits

5

Voltage converter (PT ratio)

1 to 9999.999

Current converter (CT ratio)

1 to 9999.999

Measurement items

Voltage, current, active power, apparent power, reactive power, power factor, phase angle, frequency, integrated current, integrated power, positive integrated power, negative integrated power, integration time, voltage crest factor, current crest factor, voltage peak, current peak, Thd

Displayed measurement parameters

Vdc, Vrms, V+pk, V-pk, Idc, Irms, I+pk, I-pk, P, P+pk, P-pk, VA, VAR, PF, CFV, CFI, DEG, VHz, IHz, THDV, THDI

 Voltage Measurement

Measurement range

CF=3 :

15V, 30V, 60V, 150V, 300V, 600V

CF=6 :

7.5V, 15V, 30V, 75V, 150V, 300V

Crest factor

3, 6

Accuracy

Effective range

1 % to 105 % of range

DC

±(0.2 % reading  +  0.2 % range)

45 Hz ≤ f ≤ 66 Hz

±(0.1 % reading  +  0.1 % range)

66 Hz < f ≤ 1kHz

±(0.1 % reading  +  0.2 % range)

1 kHz < f ≤ 6kHz

± 3 % of range

The filter is turned on

Increase 0.3 % reading@ 45Hz to 66Hz

Temperature effect

5-18˚C / 28-40˚C

Increase ±0.03% reading /˚C

Residual noise

0.5 % of range

 Current Measurement

Measurement range

CF=3 :

5mA, 10mA, 20mA, 50mA, 100mA, 200mA, 500mA, 1A, 2A, 5A, 10A, 20A

CF=6 :

2.5mA, 5mA, 10mA, 25mA, 50mA, 100mA, 250mA, 0.5A, 1A, 2.5A, 5A, 10A

Crest factor

3, 6

Accuracy

Effective range

1 % to 105 % of range

DC

±(0.2 % reading  +  0.2 % range)

45 Hz ≤ f ≤ 66 Hz

±(0.1 % reading  +  0.1 % range)

66 Hz < f ≤ 1kHz

±(0.1 % reading  +  0.2 % range)

1 kHz < f ≤ 6kHz

± 3 % of range

The filter is turned on

Increase 0.3 % reading@ 45Hz to 66Hz

Temperature effect

5-18˚C / 28-40˚C

Increase ±0.03% reading /˚C

Residual noise

0.5 % of range

 Power Measurement 

Accuracy

Effective range

1 % to 110 % of range

DC

±(0.2 % reading  +  0.2 % range)

45 Hz ≤ f ≤ 66 Hz

±(0.1 % reading  +  0.1 % range)

66 Hz < f ≤ 1kHz

±(0.1 % reading  +  0.3 % range)

1 kHz < f ≤ 6kHz

± 3 % of range

The filter is turned on

Increase 0.3 % reading@ 45Hz to 66Hz

Temperature effect

5-18˚C / 28-40˚C

Increase ±0.03% reading /˚C

 Frequency Measurement 

Measurement range

The filter is turned on

30.000Hz to 499.99Hz

The filter is turned off

30.000Hz to 9.9999kHz

Measurement items

Voltage, Current

Effective input range

10% to 105% of voltage input range

Accuracy

±(0.06 % reading)

Integrator Measurement 

Integrator

Accuracy

±(Accuracy of voltage or current+ 0.1 % reading)

Time

Range

0 hour 0 minute to 9999 hours 59 minutes

Accuracy

±0.01% ±1second

Subscribe now